Trace gas inlet flow & leak check
Purpose: This
procedure insures that the trace gas system (NO, NO2, CO, SO2, NOy, NOy*) is
leak tight and recording the proper flow rate.
It is very important that the gas MFCs are calibrated properly before
these tests are completed.
Required Equipment: BIOS with low cell, BIOS
tubing, four 9/16” caps,
Reference: SOPMAN
1. Down the gas channels (and met channels if on the same
tower) from set flags menu
2. Take down the tower
3. Remove inlet cones
4. Cap inlets with 9/16” caps
5. Record flow rate for CO MFC, SO2 MFC, NO RXN, NO BYP, NOy RXN and NOy BYP MFCs into
site log and in the Leak check form in the Tools drop down menu
6. Remove vacuum from the MFCs and record the flow rate
into the site log and in the Leak check form in the Tools drop down menu. This is the Pump Off number. Note: You have to remove the vacuum from both of the
MFCs that are connected to a single inlet.
Ex.: NOy
RXN and NOy BYP are on the same inlet so the vacuum
has to be removed for both MFCs to get a good Pump Off
number (NO RXN & NO BYP are connected to the same inlet. The CO MFC & SO2 MFC are connected to the
same inlet).
7. If the difference is +/- 3%, check for loose
connections and repeat check
8. Remove caps from inlet
9. Replace vacuum to MFCs
10. Connect BIOS to CO/SO2 inlet
11. Record average of three 10-step Save readings in site
log and in the Flow Audit form in the Tools drop down menu
12. Record the 1 minute average of the CO and SO2 MFCs in
the site log and in the Flow Audit form in the Tools drop down menu
13. If the difference is +/- 3%, check for loose
connections and repeat check.
14. Connect BIOS to NO/NO2 inlet
15. Record average of three 10-step Save readings in site
log and in the Flow Audit form in the Tools drop down menu
16. Record the 1 minute average of the NO and NO2 MFCs in
the site log and in the Flow Audit form in the Tools drop down menu
17. If the difference is +/- 3%, check for loose connections
and repeat check.
18. Connect BIOS to NOy/NOy* inlet
19. Record average of three 10-step Save readings in site
log and in the Flow Audit form in the Tools drop down menu
20. Record the 1 minute average of the NOy and NOy* MFCs
in the site log and in the Flow Audit form in the Tools drop down menu
21. If the difference is +/- 3%, check for loose
connections in the inlet and against the NOy moly
converter and repeat check.
22. Save all data in the Flow Audit and Leak Check forms
23. Reconnect the inlet cones to the inlets
24. Raise the tower
25. Up trace gas channels (and Met channels if down)