Tekran Flow Checks
Purpose: This procedure is in place to show how a
usual flow check is done on the Tekran when an 1130 Pump Module, 1130 Denuder
Module and an 1135 Particulate Module are connected to the 2537.
Required Equipment: BIOS with DC-2 Base
Reference: SOPMAN Pages XX-XX
Before starting the flow
checks take notice of the sampling sequence.
The Tekran must be running in Hg(0) mode to
perform flow checks.. If the Tekran is
in any other mode then wait for it to finish the cycle before initiating flow
checks.
Inlet Flow Checks
- The inlet flow check should be completed before
changing the denuder or soda lime trap.
- Remove the heated boot on the bottom of the 1130
denuder module.
- Disconnect the sample inlet at the bottom of the
zero air connector.
- Connect the flow adapter to the bottom of the
zero air connector and connect the BIOS to the adapter. Note: When connecting the adapter to the BIOS be sure to connect the tubing to the right side of the
BIOS. Also be sure to use the High
Flow Cell on the DC-2 Base.
- Take three 10 step V Averages and S Averages. Average those and then collect the
instrument information.
- Use the ` a to change the display on the 2537 to RUN.CURRENT. This menu will display the Sample
Flow rate on the 2537. It is located
on the second row on the far left and is labeled “SamFlow”
(as shown below).

- On the 1130 Pump Module locate the selector
switch for flow rate and turn it to “Current Flow”. Above the switch is the flow rate
display in LPM (Liters Per Minute).
- Write the BIOS Volumetric and STP flow rates,
2537 flow rate and 1130 Pump module flow rate in your site log as STEP 1 of the Tekran flow checks. Also record the date and time of the
flow check.
- Install the new Quartz Denuder, replace the sample
inlet and heated boot as shown in the “Tekran Denuder Changeout”
procedure.
Analyzer Flow Checks
- Make sure the analyzer is still in Hg(0) mode.
- Remove the tubing ahead of the TEE before the soda
lime trap and connect the BIOS. Note:
When connecting the adapter to the BIOS be
sure to connect the tubing to the right side of the BIOS. Also be sure to use the High Flow Cell
on the DC-2 Base.

- Take three 10 step V Averages and S
Averages. Average those and then
collect the instrument information.
- Use the ` a to change the display on the 2537 to RUN.CURRENT. This menu will display the Sample
Flow rate on the 2537.
- On the 1130 Pump Module locate the selector
switch for flow rate and turn it to “Current Flow”. Above the switch is the flow rate
display in LPM (Liters Per Minute).
- Write the BIOS Volumetric and STP flow rates,
2537 flow rate and 1130 Pump module flow rate in your site log as STEP 2 of the Tekran flow checks.
- Reconnect the tubing in front of the TEE before
the Soda Lime trap.
- Remove the tubing ahead of the soda lime trap and
connect the BIOS. Note:
When connecting the adapter to the BIOS be
sure to connect the tubing to the right side of the BIOS. Also be sure to use the High Flow Cell
on the DC-2 Base.

- Use the ` a to change the display on the 2537 to RUN.CURRENT. This menu will display the Sample
Flow rate on the 2537.
- Write the BIOS Volumetric and STP flow rates and
2537 flow rate in your site logs as STEP
3 of the Tekran flow checks.
- Changeout the Soda Lime Trap as shown in the
“Tekran Soda Lime and Inlet Frit Changeout” procedure.
- All of the inlet flow checks are now complete.
All of the
information written into the site log for STEP 1, STEP 2 and STEP 3 should now
be typed into the SM ticket.
When all flow checks
and maintenance have been completed be sure to put the Tekran back into RUN
mode.