Sunset Laboratory OCEC Sample Flow Adjustment
Purpose: To set the inlet flow rate to 8
lpm (volumetric – not STP)
Required Tools: Bios – High Cell
Overview: If the Sunset Sample flow is outside 8.0 +/- 0.25 lpm
it should be adjusted back to 8.0. This
is accomplished by adjusting a needle valve inside the instrument itself. The sample flow adjustment should be carried
out during a scheduled filter change if needed.
Install BIOS and measure/adjust sample flow
1.
Disconnect
the inlet tube from the Sample port on the Sunset instrument and attach the
BIOS with the High cell installed.
2.
Open the
“Filter installation & access port” door,
3.
In the
Sunset “Semi-continuous Real-time Carbon Analysis” window, in the top center,
click (to uncheck) the “Use Sample File Times”
4.
Then
click “Start Analysis” in the upper right hand corner of the window.
5.
You
should hear the “Ball Valve” operate and approximately “8.00” should display in
the “Sample Flow L/min” field of the “RT1 GAS FLOWS AND ABSOLUTE PRESSURE”
window, indicating that the sample flow is on.
6. If the Sample flow is outside 8.00 +/- 0.25 lpm adjust the knurled knob approx 3/8" in diameter
that you turn to open or close the valve and set it to 8.00.
7. After the flow is set reattach the inlet
tubing and check the flow at the inlet.
Flow adjusting valve located in this area
![]()

Restoring analyzer
to normal operation
1.
In the
Sunset “Semi-continuous Real-time Carbon Analysis” window, Click (to check) the
“Use Sample File Times”
2.
Click
“Start Analysis”
3.
In the
“RT1 Status” window the “Start Collection Time” field will display the start
time of the next sample, usually on the hour, every hour.
Finish Up
1.
In TrackOps record activity & close the ticket.
2.
In the SiteLog record activity.