Sunset Laboratory
OCEC Filter Replacement (Rev: 1.3)
Purpose: To remove and replace the quartz sample
filter & leak check the inlet.
Required Tools: Sunset filter removal tool
Sunset filter punch
Sunset flat forceps or filter
tweezers
Inlet cap
Required Supplies:
37mm baked Quartz filter (same
as used for PCM3)
Latex/vinyl gloves
Reference: “Initial Installation of Sunset Lab OCEC
Field Instrument”, “Installation Notes” section 2, part h “Filter
Installation/removal”, page 4. “Install-RT3.ppt” PowerPoint presentation. RealTimeOCECManualv8.
Overview:
The Sunset control software (RTOCEC888) will be stopped, the filter glassware
removed & the filter replaced, the software restarted and the inlet capped
to check for leaks, the oven cleaned, and finally the software set to normal
sample collection mode.
Stopping the Software:
1.
To prevent excess
data loss only stop the analyzer just after a new sample has started - samples
start on the hour, so 1 to 15 minutes after the hour is ideal.
2.
In the Sunset
“Semi-continuous Real-time Carbon Analysis” window, in the upper right hand
corner, click the “Cancel this run” button.
3.
The “VERIFY
4.
The software is
now stopped.
5.
Reboot the
computer to clear all operating system buffers.
Filter Removal and Replacement:
1.
Open the “Filter
installation & access port” door,
Filter access cover

2.
Remove the
“Filter access cover” by pulling up on the locking knob and sliding the cover
out.

3.
To gain access to the filter:
a)
First loosen the
“gland nut” and slide it back.
b)
Then loosen the
Nylon nut holding the photodetector and slide the photodetector & Nylon nut
off the quartz tube.
Stainless steel nut,
called “gland nut” Nylon fitting photodetector

Note: Do not touch the quartz insert tube
with your fingers - ware gloves. Quartz insert tube Nylon
locking nut. Quartz insert
tube Note o-ring

c)
Use gloves
whenever handling the quartz insert tube.
d)
Loosen (do not
remove) the quartz insert tube Nylon locking nut and carefully slide out the
quartz insert tube.
Note: The slotted portion of
the tube goes into the oven tube.

![]()

4.
Get the filter
removal tool and remove the protective rubber cap from the “harpoon” tip.
5.
Carefully insert
the harpoon tip end of the tool into the oven quartz tube until it just touches
the filters.
6.
Now give the tool
a quick jab so that the harpoon tip punches through the filters (there are two
filters).
7.
Carefully pull
the tool out along with the two harpooned filters. In the unlikely event that the filters cannot
be harpooned and they get pushed past the shoulder in the oven quartz tube,
they will have to be “fished” out with the hooked end of the filter tool or a
hooked tipped wire - they cannot be left in the inner oven cavity.
Caution: When “fishing” for
the filters, be careful not to damage the oven thermocouple.
Quartz
filters after
removal from
the internal oven
seat

8.
Replace the
protective cap back on the harpoon tip of the filter tool and put it away.
9.
Get out the
filter cutting tool.
10.
You will cut two
filters from one 37mm PCM3 baked quartz filter.
11.
Place the 37mm filter on a piece of aluminum foil placed on top of a firm,
but not hard, pad - a Post-it pad works fine.
Note: Do not cut the filters
on a hard surface, like a counter top; it will damage the cutting edge of the
tool.
12.
Put the filter
punch on the filter near the edge, but not over the edge, and press down firmly
to cut through the filter and not the aluminum foil.
13.
Leave the cut
filter in the tool and place the tool on the uncut area of the 37mm filter and
cut out the second filter.
Quartz
filters and filter punch

14.
Carefully remove
the filters from the punch and, using tweezers, place them in the opening of
the oven quartz tube.
New
quartz filters
ready for insertion using the
quartz insert tube

15.
Position the
filters so that they are vertical in the oven quartz tube.
16.
Put on a clean
pair of gloves and wipe the quartz insert tube to remove any finger prints,
lint, filter fragments, etc.
17.
Carefully insert
the quartz insert tube (slotted portion goes into oven) and push the filters in
until they are firmly seated against the oven shoulder.
Quartz insert tube Nylon
locking nut. Filters
vertical and being pushed into place with the quartz insert tube.

18.
While holding the
filters firmly in place against the oven shoulder, tighten the Quartz insert tube Nylon locking nut.
IMPORTANT: Hand
tighten only. Do Not use any tools
to tighten any of the nuts.
19.
Reinstall the
photodetector w/Nylon nut back on the quartz insert tube - do not tighten.
20.
Attach the
“gland” nut to the photodetector and hand tighten - make sure the o-ring is in
place.
21.
Let the
photodetector align itself on the quartz insert tube and then hand tighten the Quartz insert tube Nylon locking nut.

22.
Carefully make
sure that all nuts are firmly hand tight.
23.
Re-install the
“Filter access cover”.
24.
Close and secure
the “Filter installation & access port” door.
Leak Check:
1.
In the Sunset
“Semi-continuous Real-time Carbon Analysis” window, in the top center, click
(to uncheck) the “Use Sample File Times”
2.
Then click “Start
Analysis” in the upper right hand corner of the window.
3.
You should hear
the “Ball Valve” operate and approximately “8.00” should display in the “Sample Flow
L/min” field of the “RT1
4.
Cap the inlet.
5.
The sample flow
should drop to approximately .16LPM or lower (2% of 8 LPM).
6.
If not, move the
inlet cap to the “Sample Inlet” port on the back of the analyzer.
7.
If the leak check
is good, then there is a leak somewhere in the inlet, find and resolve the leak
before restoring the analyzer to normal operation.
8.
If the leak check
is still bad, then there is a leak somewhere in the sample circuit inside the
analyzer.
9.
Re-check all of the
tubing connections for tightness.
CAUTION: Hand
tighten only - do not use any tools to tighten any of the connections.
10.
Make
sure all of the plastic ferrules and o-ring are in place.
11.
If
you cannot resolve the leak, contact
12.
After
the system is leak tight, remove the cap, and restore all inlet connections.
Clean the Ovens:
1.
In the Sunset
“Semi-continuous Real-time Carbon Analysis” window, in the upper right hand
corner, click the “Cancel This Run” button.
2.
The “VERIFY DATA
LOSS” window will open, click the “Yes” button
3.
In the Sunset
“Semi-continuous Real-time Carbon Analysis” window, in the upper left hand
corner click on “Run”, in the drop down menu click on “Clean Oven”
4.
The ovens will
now run through a cleaning cycle that takes approximately 6 minutes.
Restoring Analyzer to
1.
In the Sunset
“Semi-continuous Real-time Carbon Analysis” window, Click (to check) the “Use
Sample File Times”
2.
Click “Start
Analysis”
3.
In the “RT1
Status” window the “Start Collection Time” field will display the start time of
the next sample, usually on the hour, every hour.
4.
The analyzer is
now set to normal operation
Finish Up:
1.
In TrackOps
record activity & close the “Sunset EC/OC Filter Change” ticket.
2.
In the SiteLog
record activity.
*********************
Procedure Complete
*****************************