Particle Concentrator Inlet Cleaning
The “Slit-Virtual Impactor” (Particle Concentrator Inlet) consist of 4 parts:
4. Minor Flow Outlet 3. Collection Nozzle 2. Acceleration Nozzle 1. “Total Flow Inlet”

Note: Before disassembling the inlet, put some alignment marks on the parts for easy reassembly.
4. Minor Flow Outlet 3. Collection Nozzle 2. Acceleration Nozzle 1. “Total Flow Inlet” Particle Concentrator disassembled

Acceleration Nozzle &
Collection Nozzle with heavy accumulation of debris

Start the cleaning with
distilled water and a ¼” wood dowel with the tip cut to a chisel point. Rub the dowel tip across
the acceleration & collection nozzle slits until all the debris has
been scrubbed off. Rinse with distilled
water.

Clean the dowel tip by
cutting a new chisel tip. On the reverse side of
the acceleration & collection nozzle slits, use the dowel tip to clean
the slit. Rinse with distilled water.

Note: All three O-rings are interchangeable.

Use a single edge razor
blade to clean the inside edges of the slits. Do not use any metal object that does not easily fit
into the slit - doing so will damage the slit. Do a final rinse by
squirting water through the slits, wipe and/or blow dry. Reassemble, reinstall and
leak check the PC system.