PCM Inlet Flow and Leak Checks
Purpose: To ensure that the PCM system is leak tight and
no external air (ie. unfiltered air) is entering the
airflow. This should be done on a day that the sampler is not running
preferably after the sample change-out has been completed. This can also be
done in conjunction with Dichot and NH3 Inlet
Flow/Leak checks.
Required Equipment: Bios with High cell
White Teflon inlet adaptor (male and female)
Pen and Site log
Reference: SOPMAN
- Down the PCM channels via the Set Flags button on
the site analyzer screen.
- Equip the Bios with the High Cell and the inlet
(male threaded) flow adapter.
- To turn the PCM on, use the “Set PCM and Dichot Start Dates” prompt found under “Setup” on the
toolbar in the ARASiteManager Screen/DAS. Note:
This triggers the Dichot, PCM and NH3 pumps.
- Change the date to today’s date by using the “drop
down” arrow and selecting ex: “Today: 4/22/2004. Hit the “Update Sample Start Dates” button.
- Obtain the “Pump Off” value for the leak check by
removing the vacuum side of the PCM MFC’s. Record this value in your site log.
- Put these lines back in place and then
record (in the site log) the flow rates from the DAS for the PCM1 and
PCM3.
- Take the BIOS up to the roof.
- Remove the PM10 Cyclone inlet/gooseneck and set
aside.
- Connect the Bios tubing from the “Outlet”
side of the Hi-Flow cell to the inlet using the inlet (male threaded) flow
adapter. Be careful that the tubing does not crimp thus restricting the
flow.
- Turn the BIOS on and wait until the Main Menu
appears. Hit the “Enter” key. Hold down both the “Select” and “Enter”
buttons on the front of the BIOS at the same time for 2-3 seconds then
release to begin recording the flow.
- Record (in the site log) three “VAve” at 10 cycles from the BIOS.
- Obtain the temperature from the Psychrometer and the BP from the DAS and record these
in the site log.
- For the leak check, you will need to turn the
pumps off by resetting the date in the “Set PCM and Dichot
Start Dates window.
- Cap the inlets using red caps.
- Turn pump back on and wait for
the flow to fall to/near zero. Please
note: If a leak can not be resolved
and the problem has been narrowed to the filterpack
itself, please mark the filterpack
and return to Mactec with a note stating “the filterpack failed leak test and should be removed from
circulation.
- Record the “Inlet Capped” value in the site log.
- While pumps are still running, carefully remove the caps from the inlets.
- Replace the PM10 Cyclone inlet/gooseneck.
- Change the “Set PCM and Dichot
Start Dates” back to the original date of 1/7/2000 and hit the “Update Sample Start Dates”.
- With the information that you have obtained,
enter the data in the electronic
form (which is found on the toolbar under Flow Checks>Flow Audit) under
the appropriate “PCM Inlet”.