Nephelometer Zero & Span Calibration
Purpose: Follow
these steps to Zero and Span calibrations of the Nephelometer instrument.
Required Equipment: Nephelometer Calibration Kit
Pure
CO2 to use for span gas
Psychrometer
Reference: SOPMAN
Notes:
Calibration is accomplished by
filling the instrument with gases of known but different scattering coefficients
and adjusting the zero and span constants for the appropriate
scattering values (bs or bscat) on either the display screen or analog output.
Particle free air is used as a zero reference scatter and one of several
gases can be used as an upscale scatter. Pure CO2 will be introduced in this
procedure to be the span reference.
Procedure
Time: Calibration time depends upon
the signal averaging time selected since that is the time required for the
output to stabilize. However, pressing the reset switch to average
periodically will decrease the stabilization time. The procedure should take
only 20 to 30 minutes. The instrument will have to be removed from sample
plumbing and set on workbench to carry out procedures. Vibration will effect
this instrument so be sure you do not bump or have it setting on a vibrating
surface.

Procedures:
Zero Calibration: Bench
Top Method
- Start by filling out the calibration
form up to the Nephelometer Zero Cal section.
- Down the Aethalometer &
Nephelometer channels before starting any manual calibration.
- Disconnect the instrument
and set up on a bench top work area.
- Vibration will effect this
instrument so be sure you do not bump or have it setting on a vibrating
surface. Picture of this setup can be found in figure #1 below.
- Turn power on to the
instrument.
- Thread the two PVC adapter
fittings from calibration kit into the inlet and outlet ports.
- Connect the zero filter assembly
on the top inlet port.
- Connect PVC fitting with
the box fan bolted onto the lower outlet port.
- Plug white electrical plug
from fan into connector supplied on back of control box.
- Allow fan to run for ten
minutes before taking readings.
- Press the Next toggle switch
three times and read the Wall % number on the instrument display. Enter
this into the calibration form. If the Wall % number is greater than 75%
contact the Plano office.
- Use the Display toggle
switch to obtain the Zero Calibration Screen as shown in picture below.

- Read the bscat number.
This number should be 0.00 to + 0.05e-5. If this number is between the specifications
your Zero is good. Set the Neph channel to DAP Integer and obtain the
number corresponding to the specification and add this to the DAP integer
line of the form. Continue to the Span calibration portion of this
procedure.
- If this number is not within
specification continue on with this calibration portion of the procedure.
- In the lower right hand
corner of the display there is a Set Zero number. Write it down in the
Set Zero as found line of the calibration form. Use the Parameter toggle
switch to adjust this number to be less or more. As this parameter number
changes watch the bscat to obtain the specification listed earlier. You
will find that as you make an adjustment you must wait for the bscat data
to settle down. Use one minute intervals to fill in the value portions of
the calibration form. Once you have reached 0.00 to +0.05e-5 the Zero
calibration is complete.
- Write into the calibration
form the final Set Zero after parameter number. Set the Neph channel to
DAP Integer and obtain the number corresponding to the specification and
add this to the DAP integer line of the form.
- Continue to the Span calibration
portion of this procedure.
Zero Calibration: Inlet Connected / Ball Valve Method
- Start by filling out the
calibration form to the Nephelometer Zero Cal section.
- Vibration will affect this
instrument so be sure to limit vibration and or bumping it while
calibration is happening.
- Down the Aethalometer &
Nephelometer channels before starting any manual calibration.
- Select SO2 Zero from the
DAS to switch the ball valve on the Aeth/ Neph zeroing box. This switch
bypasses the inlet air through a filter to simulate a zero particle
stream. Allow the SO2 Zero to run for ten minutes before taking readings
from the Nephelometer display.
- Under Display press the
Next toggle switch three times and read the Wall % number on the instrument
display. Enter this into the calibration form. If the Wall % number is
greater than 75% contact the Plano office.
- Use the Display toggle
switch to obtain the Zero Calibration Screen as shown in picture below.

- Read the bscat number.
This number should be 0.00 to + 0.05e-5.
If
this number is between the specifications your Zero is good. Set the NEPH
channel to DAP Integer and obtain the number corresponding to the specification
and add this to the DAP integer line of the form. Continue to span calibration
portion of this procedure.
If
this number is not within specification continue on with this calibration portion
of the procedure.
- In the lower right hand
corner of the display there is a Set Zero parameter number. Write it down
in the Set Zero as found line of the calibration form.
- Use the parameter toggle
switch to adjust this number to be less or more. As this parameter number
changes watch the bscat to obtain the specification listed earlier. You
will find that as you make an adjustment you must wait for the bscat data
to settle down. Use one minute intervals to fill in
the value portions of the calibration form. Once you have reached
0.00 to +0.05e-5 the Zero calibration is complete.
- Write into the calibration
form the final Set Zero after parameter number. Set the Neph channel to
DAP Integer and obtain the number corresponding to the specification and
add this to the DAP integer line of the form.
- Continue to the Span
calibration portion of this procedure.

Figure 1: Setup for Zero Calibration Bench Top Method
Span Calibration: Bench Top Method
- Pictures of this setup are shown
below in figures 2 & 3.
- Remove the zero filter assembly
from the top port if following the bench top method. Do not remove the PVC
threaded fitting if already installed, if not install it now. Replace it
with the PVC cap that has a 1/8th inch tubing fitting with 20
feet of tubing. Run the other end of tubing out the door of the shelter.
- Remove the fan from the
bottom port only is you are following the zero bench top method. Use the
other cap and PVC fittings from the kit to completely cover the bottom port.
- At this point be sure to
cap the NEPH fitting on the AETH/NEPH manifold.
- Use the Display toggle
switch to choose the Calibration Gas Screen as shown in picture below.
The ratio should read 2.61e+0 for pure CO2. If it does not read this use
the Parameter toggle switch to raise or lower the number under set R-gas
to obtain parameter. Mark this as checked on the calibration form.

- Connect the regulator to
the CO2 bottle and verify the regulator is leak proof and purged
appropriately.
- Set the regulator to Zero
PSI and the manual flow control valve is closed.
- Connect the ¼ inch copper
tubing to the pure CO2 span gas regulator.
- Connect the other end of
the tubing with the critical orifice to the purge port at the rear base of
the instrument.
- Before opening the pure CO2
main bottle valve, be sure the regulator is adjusted to zero psi. Now
open the main bottle valve to get pressure to the regulator and set the regulator
psi to five.
- Slowly open the manual flow
control valve on the regulator output to full open.
- Use the Display toggle
switch to obtain Span Calibration Screen shown in picture below. And
enter the Span parameter into the calibration form.

- Now that span gas is
passing through the instrument allow five minutes to for the bscat to
settle down.
- Use the Parameter toggle
switch to change the set span number to have the bscat and gas lines of
this display match each other as closely as possible. A tolerance of +/-
0.05 should be expected. You will find that as you make an adjustment you
must wait for the bscat data to settle down. Use one minute intervals to
fill in the value portions of the calibration form. Set the NEPH channel
to DAP Integer and obtain the number corresponding to the specification
and add this to the DAP integer line of the form.
- Fill out the span
calibration portion of the form completely.
- Once this has been
completed the instrument is ready to resume normal operation.
- Reconnect all plumbing.
- Record and calculate the
information called out in the checks after calibration section.
- Return all calibration kit
supplies into safe storage for later use.
- Up all of the Aethalometer
& Nephelometer channels.

Figure 2: Setup for Span Calibration Bench Top Method

Figure 3: Span Calibration Bench Top Method with pure CO2
bottle connected.
Nephelometer Maintenance
Cleaning Procedure for the
Scattering Chamber
- Turn power off and
disconnect instrument from sample plumbing.
- Remove the four 6-32 socket
screws from the top of the instrument. The top is sealed with an O-ring
and may require tapping to get loose from the body. Mark the housing
before fully removing it for proper baffle assembly later. The component
you are removing is called the dark trap. Set this aside.
- Using the Zero air in the
shelter blow out the scattering chamber; don’t use too much air pressure
or the chopper could become damaged. Blow out all visible debris.
- Check carefully with a
flashlight for spider webs and debris, especially around the aperture
hole. Even the smallest filament in the scattering cone can cause
extremely high scattering background.
- Now instrument is ready to
be assembled.
- Verify the O-ring is in
good condition and the mating surfaces do not have burrs on them that
could cause light to escape.
- Install the dark trap,
being careful to align the baffle with the lamp assembly.
- Install the four 6-32
socket screws onto the top of the instrument.
- Connect the sample tubing
back to instrument. Turn the power back on.
Changing the Flash Lamp
Assembly
Caution – The 2.2uf
capacitor inside the flash lamp assembly stores approximately 400vdc, so
disconnect power before removing cover.
- Remove the two 8-32 screws
securing the flash assembly cover and remove it.
- Remove the four 4-40 socket
screws holding the lamp assembly and remove. You may have to pry the
assembly away from the gasket carefully with a screw driver.
- The lamp is connected to
the flash circuit board with a six pin Amp connector.
- Unplug the Amp connector
and replace lamp with new assembly.
- Once the new assembly has
been connected reverse the order to replace the lamp assembly.
Nephelometer Analog
Calibration
Analog
calibrations for this instrument are accomplished by using Zero, One and Five volts
DC. At the site the most effective way to do this is to use any C class
analyzer as a voltage generator. On the C class analyzer choose set outputs to
Zero, those analyzers analog output green connector voltage changes to 0 volt
DC. Choose Full Scale, the green connector voltage changes to 5 volts DC. To
obtain 1 volt DC for TP and RH cal choose Full Scale and scroll to 20% by using
the arrow keys. Use this to generate the voltages needed for the two point
analog calibration.
There
are six channels that will or may require calibration. They are named below.
Cal Channel Values and
Voltages for each are:
Zero
Value Voltage Span Value Voltage
NEPH 0 0 2e-4 5
NEPH LOG 0 0 1e-3 5
NEPH IC 0 0 1e-3 5
NEPH LB 3 0 5 5
NEPH TP -10 0 60 1
NEPH RH 0 0 100 1