Nephelometer Calibration Form
Nephelometer information before disconnection or calibration:
Site Name: ________ Site ID: ________ Date: ________
Operator: ________ Nephelometer Serial Number: ________________
Nephelometer display value (bs): ________________
Data View value (NEPH): ________________
Percent Difference of (Bs) versus (NEPH): ________________
Psychrometer Temperature at Nephelometer: ________________
Barometric Pressure from DAS (BP): ________________
Nephelometer temp from DAS (NEPHTP): ________________
Nephelometer display values:     P: ______________ T: ______________ Rh:_____________
Nephelometer Zero Calibration data:
Wall Percentage value as found: __________ (If found to be more than 75% call Plano office)
Bscat value as found:  Value 1 __________ Value 2 __________ Value 3 __________
(Bscat number should be 0.00 to + 0.05e-5)
Set Zero parameter number as found: _______________
Bscat value as left:  Value 1 __________ Value 2 __________ Value 3 __________
DAP integer from DAS (NEPH):________________
Set Zero parameter number as left:________________
Nephelometer Span Calibration data:
The ratio should read 2.61e+0 for pure CO2.  Is this value set (Y/N): __________
Bscat value as found:  Value 1 __________ Value 2 __________ Value 3 __________
Gas constant value:  Value 1 __________ Value 2 __________ Value 3 __________
(Bscat number should be as close to Gas number as possible with a +/- .05 tolerance)
Set Span parameter number as found: _______________
Bscat value as left:  Value 1 __________ Value 2 __________  Value 3 __________
Gas constant value:  Value 1 __________ Value 2 __________ Value 3 __________
DAP integer from DAS (NEPH): __________
Set Span parameter number as left: _______________
Nephelometer checks after Calibration:
Nephelometer display value (bs): ________________
Data View value (NEPH): ________________
Percent Difference Bs vs, NEPH: _______________ (Greater than +/-3% requires analog calibration)
Nephelometer display values:     P: ______________ T: ______________ Rh:_____________
Psychrometer Temperature at Nephelometer: ________________
Barometric Pressure DAS (BP): ________________
Nephelometer temp from DAS (NEPHTP): ________________
CO2 bottle pressure remaining (psig): ________________
Wall Percentage value: _______________ (If found to be more than 75% call Plano office)
Flow Check:________________(4-6SLPM) Leak Check: ________________
Did instrument require cleaning:________________
Signed off by:________________
Date:________________