| Nephelometer Calibration Form | ||||||||
| Nephelometer information before disconnection or calibration: | ||||||||
| Site Name: | ________ | Site ID: | ________ | Date: | ________ | |||
| Operator: | ________ | Nephelometer Serial Number: | ________________ | |||||
| Nephelometer display value (bs): | ________________ | |||||||
| Data View value (NEPH): | ________________ | |||||||
| Percent Difference of (Bs) versus (NEPH): | ________________ | |||||||
| Psychrometer Temperature at Nephelometer: | ________________ | |||||||
| Barometric Pressure from DAS (BP): | ________________ | |||||||
| Nephelometer temp from DAS (NEPHTP): | ________________ | |||||||
| Nephelometer display values: | P: ______________ | T: ______________ | Rh:_____________ | |||||
| Nephelometer Zero Calibration data: | ||||||||
| Wall Percentage value as found: __________ (If found to be more than 75% call Plano office) | ||||||||
| Bscat value as found: | Value 1 __________ | Value 2 __________ | Value 3 __________ | |||||
| (Bscat number should be 0.00 to + 0.05e-5) | ||||||||
| Set Zero parameter number as found: _______________ | ||||||||
| Bscat value as left: | Value 1 __________ | Value 2 __________ | Value 3 __________ | |||||
| DAP integer from DAS (NEPH):________________ | ||||||||
| Set Zero parameter number as left:________________ | ||||||||
| Nephelometer Span Calibration data: | ||||||||
| The ratio should read 2.61e+0 for pure CO2. Is this value set (Y/N): __________ | ||||||||
| Bscat value as found: | Value 1 __________ | Value 2 __________ | Value 3 __________ | |||||
| Gas constant value: | Value 1 __________ | Value 2 __________ | Value 3 __________ | |||||
| (Bscat number should be as close to Gas number as possible with a +/- .05 tolerance) | ||||||||
| Set Span parameter number as found: _______________ | ||||||||
| Bscat value as left: | Value 1 __________ | Value 2 __________ | Value 3 __________ | |||||
| Gas constant value: | Value 1 __________ | Value 2 __________ | Value 3 __________ | |||||
| DAP integer from DAS (NEPH): __________ | ||||||||
| Set Span parameter number as left: _______________ | ||||||||
| Nephelometer checks after Calibration: | ||||||||
| Nephelometer display value (bs): | ________________ | |||||||
| Data View value (NEPH): | ________________ | |||||||
| Percent Difference Bs vs, NEPH: _______________ (Greater than +/-3% requires analog calibration) | ||||||||
| Nephelometer display values: | P: ______________ | T: ______________ | Rh:_____________ | |||||
| Psychrometer Temperature at Nephelometer: | ________________ | |||||||
| Barometric Pressure DAS (BP): | ________________ | |||||||
| Nephelometer temp from DAS (NEPHTP): | ________________ | |||||||
| CO2 bottle pressure remaining (psig): | ________________ | |||||||
| Wall Percentage value: _______________ (If found to be more than 75% call Plano office) | ||||||||
| Flow Check:________________(4-6SLPM) | Leak Check: | ________________ | ||||||
| Did instrument require cleaning:________________ | ||||||||
| Signed off by:________________ | ||||||||
| Date:________________ | ||||||||