NO2 Converter – Check/ Add Cooling Water
Purpose: This assures that the cooling water for the
NO2 Power Supply is always 3/4 full.
Required Equipment:
DI
Water or DI Water with an Anti-Fungal Mixture
NO2 Power Supply Water
Bottle
One NO2 power supply
quick disconnect fitting
Reference: SOPMAN
- Locate the NO2 power supply water reservoir.
- Observe the water level. If it is not at least 3/4 full, add
water.
- To add water to the reservoir use the NO2 power
supply water bottle. It has a quick
disconnect fitting on the end that is made for the water ports on the
front of the power supply.
- Insert the extra quick disconnect fitting in the
“Vent” port on the left. This so
that air can escape as you fill the reservoir.
- Insert the water bottle with the quick disconnect
fitting in the “Fill” port on the right and squeeze the bottle so that
water will enter the reservoir.
- Once the reservoir is at least 3/4 full remove
the two quick disconnect fittings and close the scheduled maintenance
ticket.