Met Sensor – Precipitation 473ml 100-Tip Audit
Purpose: This should be done along with all other Met Sensor
checks.
Required Equipment: 500 ml measuring cylinder
Plastic container that is used for the PRECIP check
Reference: SOPMAN
- Down the channel for PRECIP
on the SiteManager screen via “Set Flags” button.
- Locate the 500:5ml
measuring cylinder and add 473ml of water. Pour this into the plastic
container that is used for the PRECIP check. (Make sure the valve is
closed before adding water.)
- Take the screen out of the
PRECIP bucket.
- While holding the container
of water over the PRECIP bucket, slowly open the valve until you have a slow,
steady flow of water. (Note: the rocker needs to tip about 4-5 times a
minute.)
- Open the “Query Data”
located under “Tools” on the toolbar.
- Select PRECIP from the drop
down box, and hit the “Get Data” button to update. (It may be necessary to
wait until the last minute average has updated.)
- Wait until all the water
has left the container and update the data. Add the values for this time.
This should add up to a total value of 25.4 (If this value is not correct,
it may be necessary to adjust the rocker arm by the two small screws
located under the PRECIP bucket. Call Plano for instructions.)
- Up the PRECIP channel and
record your results in the “MetChecks” form found under “StatusForms” on
the tool bar.