Met Sensor - Precipitation 10-Tip Audit
Purpose: This should be done along with all other Met Sensor
Checks.
Required Equipment: (none)
Reference: SOPMAN
- Down the channel for PRECIP
on the SiteManager screen via “Set Flags” button.
- Take the screen out and the
cover off the PRECIP bucket. You will see the rocker arm inside.
- With you finger, slowly tip
the rocker 10 times. (Waiting about 10 seconds between tips.)
- Replace the cover and
screen in the PRECIP bucket.
- Open the “Query Data”
located under “Tools” on the toolbar.
- Select PRECIP from the drop
down box, and hit the “Get Data” button to update. (It may be necessary to
wait until the last minute average has updated.)
- Add the values for the last
3-4 minutes. This should add up to a total value of 2.54. (If this value
is not correct, it may be necessary to adjust the rocker arm by the two
small screws located under the PRECIP bucket. Call Plano for
instructions.)
- Up the PRECIP channel and
record your results in the “MetChecks” form found under “StatusForms” on
the tool bar.