Fixed and Variable Set Point MFC
Audit Procedure
Purpose: This procedure is to show the steps for Auditing
a Fixed or a Variable Mass Flow Controller (MFC). If the Fixed MFC audit error is greater than
+/- 2% it will need to be calibrated. If
the Variable Set Point audit error is greater than +/- 1% it will need to be
calibrated. The error is based on the
DAS read out for the MFC and the S avg on the BIOS.
Required Equipment: BIOS
BIOS
flow cells (Low, Standard or High)
Flow
tubing for BIOS
Wrenches
(7/16, 9/16 or 11/16)
DAS
(DataView, Query Results, Cal Channel and/or Trigger Cals)
Reference: SOPMAN Pages XX-XX
MFC Flow Rates
The following table shows the
correct set points for each MFC.
MFC Setup Table
MFC Flow
Rate BIOS Cell
Used Process Control
PCM1 – 4 MFC 16.7
slpm High
Flow Cell PCM DAY1
NO RXN MFC 1.00
slpm STND
Flow Cell NO RXN FLOW
NO
BYP MFC 1.00 slpm STND Flow Cell NO BYP FLOW
NOy
RXN MFC 1.25 slpm STND Flow Cell NOy RXN FLOW
NOy
BYP MFC 1.25 slpm STND Flow Cell NOy BYP FLOW
NO3CH1MFC 1.25 slpm STND Flow Cell NO3 CH1 FLOW
NO3CH2MFC 1.25 slpm STND Flow Cell NO3 CH2 FLOW
NO3CH3MFC 1.25 slpm STND Flow Cell NO3 CH3 FLOW
SO4
MFC 1.25 slpm STND Flow Cell SO4 FLOW
CO
MFC 1.00 and
1.25 slpm STND Flow Cell CO FLOW
SO2
MFC 1.00 and
1.25 slpm STND Flow Cell SO2 FLOW
NH3
MFC 16.7 slpm High Flow Cell DICHOT/NH3 ON
Dichot
Coarse MFC 1.67 slpm STND Flow Cell DICHOT/NH3 ON
Dichot
Fine MFC 15.0 slpm High Flow Cell DICHOT/NH3 ON
Air
MFC Variable High Flow Cell Air MFC Cal
Gas
MFC Variable Low Flow Cell Gas MFC Cal
NO3
Gas MFC Variable Low Flow Cell NO3 Gas MFC Cal
SO4
ADD MFC Approx. 5 SCCM Low Flow Cell SO4 ADD
Using the DC-2 BIOS
- Turn
on the BIOS. It will go through an
initial start up and come to a screen with options RUN, SETUP and LEAK
CHECK.
- At
this point the cursor will be blinking on the RUN option. Press enter to continue to the next
screen. DO NOT go into the setup
mode.
- To
obtain the temperature and pressure from this screen hold the Select
button down for 3 seconds. The
display will now show the date, time, temperature, pressure and volume.
- To go
back to the flow screen push the reset button.
- The BIOS
can now be used to obtain flow rates by pressing the Select and Enter
button simultaneously.
- When
changing BIOS cells it is not necessary to turn the unit off. Simply press the reset button and the
Base unit will detect a different cell.
It is now ready to make flow readings.
Auditing a Fixed
Setpoint MFC
- Start the process by opening the Flow Audit menu on
DataView. It is located on DataView
under the Flow Checks tab.
- With the Flow Audit menu open select the MFC that is
being audited (Be sure to select the MFC and not the inlet). Now push the button labeled: “Enter
Flow Check Data”.
- Select the appropriate BIOS Cell, according to the MFC Setup Table from above, and
connect it to the MFC. If the MFC
has vacuum as the supply then connect it to the right side of the BIOS
flow cell and if the supply is pressure connect it to the left side of the
BIOS.
- If the MFC audited is PCM 1-4 open the Trigger Cals
menu and select PCM DAY 1 then push the start calibration button. The MFC will now start flowing.
- If the MFC audited is NH3 or a Dichot MFC open the
Trigger Cals menu and select Dichot/NH3 ON then push the start calibration
button. The MFC will now start
flowing.
- For Fixed Set Point MFCs turn on the BIOS
and collect 3 10 - point BIOS V averages and S averages at the MFC. Average each of those and enter them
into the MFC Flow Audit Entry Form.
- Collect a minimum of 2-1 minute averages of DAS flow
data using the Query Results tool to get the average. Enter this information into the MFC Flow
Audit Entry Form in the blank labeled ‘SiteManager Flow Rate’.
- With the MFC Flow Audit Entry Form complete push the
‘Calculate’ button. If all of the
data is correct push the ‘Yes’ button under Save Data. If not push the ‘No’ button and correct
the data.
- If the MFC Flow Error is more than +/- 2% then
calibrate the MFC. To change to
another MFC after an audit has been saved push the ‘Yes’ button under
Change Device.
Auditing a Variable Set Point MFC
- Start the process by opening the Flow Audit menu on
Dataview. It is located on DataView
under the Flow Checks tab.
- With the Flow Audit menu open select the MFC that is
being audited (Be sure to select the MFC and not the inlet). Now push the button labeled: “Enter Flow Check Data”.
- Select the appropriate BIOS Cell, according to the MFC Setup Table from above, and
connect it to the MFC. If the MFC
has vacuum as the supply then connect it to the right side of the BIOS
flow cell and if the supply is pressure connect it to the left side of the
BIOS.
- For a Variable Set Point MFC open the Cal
Channel window by selecting “Calibrate DAS Channel” under the Dataview
Setup tab. With Cal Channel open
select the correct channel name of the MFC being calibrated. The DAS value and instantaneous DAP
Integer will appear under the channel name.
- The cal channel window should now have the correct
MFC selected under channel name.
Each MFC has a process control associated with it. Using the MFC Setup Table from above select the appropriate process
control in the “Process Control List” at the bottom of the Cal Channel
window. Then press the Run Selected
Process Control button.
- Now that the MFC is selected, input 5000 into the
“New DAP Value” window and push the “Update DAP Value” button.
- Collect 3 10 - point BIOS V averages and S averages
at the MFC. Average each of those
and enter them into the MFC Flow Audit Entry Form.
- Collect a minimum of 2-1 minute averages of DAS flow
data using the Query Results tool to get the average. Enter this information into the MFC Flow
Audit Entry Form in the blank labeled ‘SiteManager Flow Rate’.
- With the MFC Flow Audit Entry Form complete push the
“Calculate” button. If all of the
data is correct push the “Yes” button under Save Data. If not push the “No” button and correct
the data. For each setpoint note the MFC Flow Error percentage.
- Repeat steps 7-9 for a DAP value of 10000, 15000,
20000 and 30000.
- If the MFC Flow Error on any of the setpoints is more
than +/- 1% then calibrate the MFC.
To change to another MFC after an audit has been saved push the
“Yes” button under Change Device.