Hi-Vol Motor Replacement and Calibration
Purpose: The procedure below is to give explanation on how to
replace a Hi-Vol motor. When replacing a motor it will have to be calibrated
before going back in service.
Items needed:
New motor
Crescent wrench
Medium phillips head and
flat blade screwdriver
Calibration orifice
Monometer
Reference: Tisch Environmental Hi-Vol Operations Manual pages
6-17.
Motor change out
- Unplug the machine.
- Open up the front and
disconnect the black flow tubing from the chart recorder.
- Disconnect the power cord
for the motor and take the exhaust tube off of the motor housing.
- Open the top and take the
filter and filter plate out.
- Pull up on the filter
screen and the motor will come with it. Make sure that you take the flow
sensor off as soon as you are able to reach it.
- You should be able to pull
the motor housing and filter screen out now. It is one complete piece.
- After pulling it out,
disconnect the filter screen plate from the motor housing. There are four
screws that hold it together. When the two plates separate there will be
a rubber gasket. Make sure you save it to use for reinstallation. Be
careful, with the two plates apart the motor will fall out.
- Loosen the knurled nut that
holds the power cord in place and push the cord through as far as
possible.
- Now you can pull the motor
from the housing.
- Disconnect the wires and
replace the motor. Make sure you take the rubber seal and aluminum plate
off of the old motor before you connect the wiring.
- Install the new motor, then
its housing and then the filter screen.
- When everything is back in
place connect the power cord for the motor and plug the main unit in. DO
NOT put the black flow tube back on to the chart recorder.
Calibration of new motor
- Put the flow orifice in
place where the filter goes and hang your monometer on the side of the
Hi-Vol housing.
- Turn the orifice to the
fully open position and turn the “timed/off/on” switch to “on” so that the
Hi-Vol starts running.
- Connect the tubing on the
orifice to the monometer. Connecting the tubing to either side will give
you the same results. Which ever side you choose, make sure you use it
for the entire calibration.
- The monometer has liquid in
it that will rise and fall on each side depending on the flow.
- Adjust the flow orifice
using the valve on the top side so that the two sides of the monometer
added together equal 9.5.
- Write the value from each
side of the monometer on to the Hi-Vol calibration form under the
“Orifice” heading. The form is attached to the bottom of this procedure.
- Pull the orifice tubing off
of the monometer and put the black tubing from the motor housing onto the
monometer.
- Write the value from each
side of the monometer on to the Hi-Vol calibration form under the “Motor”
heading.
- Repeat steps 18 – 21 for
set points 8.0, 6.75, 4.5 and 2.0.
- The cal form should have
all five points filled in for both the Orifice and Motor. Make sure you
input the values for both the Left and Right side of the monometer.
- Write the current pressure
and temperature on to the Hi-Vol calibration form.
- Fax a copy of the motor
calibration form to Plano.
- Push the motor housing
tubing back on to the chart recorder and replace the chart.
- Install a new filter.
- Check the time and adjust
it if needed.
- Set up the machine for the
next run date.
- Reset the run time so that
it reads 0.0.
- After the calibration and
set up is complete put the “start day setting” in the correct position and
put the “timed/off/on” switch into the “timed” position.
Example calibration form:

The blank calibration form does have calculations imbedded
in it.
