Dichot Inlet Flow and Leak Check
Purpose: The Dichot Inlet flow is split into two flows that
measure on the DiCoarse filter and DiFine filter. (You will see these on the
ARASiteManager screen.) An inlet flow is compared against these two flow
measurements on the DAS.
Required Equipment: Bios with High Cell
Psychrometer
Inlet
flow adapter (same as 5400 and FRM)
Reference: SOPMAN
- This should be done on a
day that the sampler is not running preferably after the sample change-out
has been completed. This can also be done in conjunction with PCM and NH3
Inlet Flow/Leak checks.
- Down the DICOARSE and
DIFINE channels via the Set Flags button on the site analyzer screen.
- Equip the Bios with the
High Cell and the inline flow adapter used with the Dichot and FRM.
- To turn the Dichot on, use
the “Set PCM and Dichot Start Dates” prompt found under “Setup” on the
toolbar in the ARASiteManager Screen/DAS. Note: This triggers the Dichot,
PCM and NH3 pumps.
- Change the date to today’s
date by using the “drop down” arrow and selecting ex: “Today: 4/22/2004. Hit the
“Update Sample Start Dates” button.
- Obtain the “Pump Off” value
for the leak check by removing the vacuum side of the Dichot MFC’s. Record
this value in your site log.
- Put these lines back in
place and then record (in the site log) the flow rates from the DAS for
the DiFine and DiCoarse.
- Take the BIOS up to the
roof.
- Remove the inlet and set
aside.
- Connect the Bios tubing
from the “Outlet” side of the Hi-Flow cell to the inlet using the inlet
flow adapter. Be careful that the tubing does not become crimped thus
restricting the flow.
- Turn the BIOS on and wait
until the Main Menu appears. Hit the “Enter” key. Hold down both the
“Select” and “Enter” buttons on the front of the BIOS at the same time for
2-3 seconds then release to begin recording the flow.
- Record (in the site log) three
“VAve” at 10 cycles from the BIOS.
- Obtain the temperature from
the Psychrometer and the BP from the DAS and record these in the site log.
- For the leak check, close
the flow audit adapter and carefully remove the tubing.
- Get the leak check
information from the DAS when the flow drops down to/near zero. This is
considered the “Inlet Capped” value of the leak check.
- Slowly turn the flow audit
adapter until the flow is resumed and take the adapter off.
- Replace the inlet.
- Change the “Set PCM and
Dichot Start Dates” back to the original date of 1/7/2000 and hit the “Update
Sample Start Dates”.
- With the information that
you have obtained, enter the data in the electronic form (which is found
on the toolbar under Flow Checks>Flow Audit) under “Dichot Inlet”.