TPN Audit and Calibration
Purpose: The TPN audit and calibration is to assure
that the TPN analyzer/system has the correct gas response. This can be accomplished by introducing a
known quantity of gas to the inlet and auditing the response on the instrument. The terminology for this type of audit is
Required Equipment: Precision flat head
screwdriver
Analyzer Setup
1. The Background (BKG) and Coefficient (COEF) on
the analyzer should be set to 0.0 and 1.000 respectively.
2. On the “C” or “
Analyzer Audit
1. Open the Gas Calibration Worksheet and enter the correct NO
concentration (found on the gas cylinder) in the blank provided.
2. In Site Manager go to the Tools tab and select
Trigger Cals.
3. In the Trigger Cals menu select TPN Zero and
press Start Calibration.
4. Under the Tools tab select Query 1-Minute Data.
5. In the Query Data Tool menu select the Air
6. Wait 13 minutes for NO# and TPN are stable. Contact
7. Enter the Air
8. Go to Trigger Cals and press Stop Calibration
then select TPN NO
9. Wait 13 minutes for NO3 and TPN are stable then
press Get Data.
10. Enter the Air
11. The RF is now displayed in the worksheet. If the RF is between 0.98 and 1.02, audit is
complete. Close the Trigger Cals and
Query Data windows.
12. Save the Gas calibration worksheet.
13. If the RF is not within specification continue
with the Analyzer Calibration procedure.
Analyzer Calibration
1. Enter the As Found data for TPN Zero in the As
Left section for TPN Zero.
2. In Site Manager go to the Tools tab and select
Trigger Cals. In the Trigger Cals menu
select TPN Zero and wait 5 minutes for data to be captured for First Light use
3. In the Trigger Cals menu select TPN NO
4. Under the Tools tab select Query 1-Minute Data.
5. In the Query Data Tool menu select the Air
6. Wait 13 minutes for NO3 and TPN are stable.
7. Adjust the instrument to the Actual Expected
value from the As Found section of the worksheet.
a. On “C” or “
8. When the analyzer calibrations are complete, let
the analyzer stabilize for 5 minutes.
9. On the Query 1-Minute Data tool press Get Data.
10. Enter the Air
11. The RF is now displayed in the worksheet. If the RF is between 0.98 and 1.02, the
calibration is complete. If the RF is
not within specification, repeat the calibration part of the procedure again.
12. When the instrument is finished being calibrated
make sure to put it back into the Run mode.
13. Close the Trigger Cals and Query Data windows.
14. Save the Gas Calibration worksheet.
TPN Conversion Efficiency Audit
1.
This procedure will
obtain the current RF value for the NO2 Conversion Efficiency.
2.
Open the Gas Calibration
Worksheet and go to the NO2 Conv Eff Worksheet tab. Enter the correct NO and NO2 concentrations
(found on the gas cylinder) in the blanks provided.
3.
An NO audit of the
analyzer has to be completed before the NO2 conversion efficiency can be
completed. You may use that day’s
information if available. If it is
available enter the information and start with Step 15. Otherwise continue with the next Step.
4.
On the Site Manager
screen go to the Tools tab and select Trigger Cals.
5.
In the Trigger Cals menu
select TPN Zero and press Start Calibration.
6.
Under the Tools tab
select Query 1-Minute Data.
7.
In the query data tool
menu select the Air MFC, Gas MFC, NO3 and TPN.
8.
Wait 13 minutes and
verify the Zeros for NO3 and TPN are stable (+/-0.1 ppb). Contact
9.
Enter the Air MFC, NO3
and TPN data in the appropriate blank for the TPN Zero on the NO2 Conv Eff
Worksheet in the TPN Analyzer Response section.
10. Go to Trigger Cals and press Stop Calibration
then select TPN NO GRC Point 1 and press Start Calibration.
11. Wait 13 minutes and verify the data for NO3 and
TPN are stable (+/- 0.2ppb), then press Get Data.
12. Enter the Air MFC, Gas MFC, NO3 and TPN data in
the appropriate blanks for the TPN NO GRC Point 1 on the NO2 Conv Eff Worksheet
in the TPN Analyzer Response section.
13. The RF is now displayed at the far right of the
worksheet.
14. Save the Gas Calibration Worksheet.
15. In the Trigger Cals menu select TPN Zero and
press Start Calibration.
16. Under the Tools tab select Query 1-Minute Data.
17. In the query data tool menu select the Air MFC,
Gas MFC, NO3 and TPN.
18. Wait 13 minutes and verify the Zeros for NO3 and
TPN are stable (+/-0.1 ppb). Contact
19. Enter the Air MFC, NO3 and TPN data in the
appropriate blank for the TPN Zero on the NO2 Conv Eff Worksheet in the TPN NO2
Conversion Efficiency section.
20. Go to Trigger Cals and press Stop Calibration
then select TPN NO2 GRC Point 1 and press Start Calibration.
21. Wait 13 minutes and verify the data for NO3 and
TPN are stable (+/- 0.2ppb), then press Get Data.
22. Enter the Air MFC, Gas MFC, NO3 and TPN data in
the appropriate blanks for the TPN NO2 GRC Point 1 on the NO2 Conv Eff
Worksheet in the TPN NO2 Conversion Efficiency section.
23. The RF is now displayed at the far right of the
worksheet.
24. If the RF is less than 0.95 it is out of spec
and the RF will turn red.
25. Contact
26. Close the Trigger Cals and Query Data windows.
27. Save the Gas Calibration Worksheet.