TPN Response Check/
Adjustment
Purpose: The TPN response check/ adjustment is to assure that the TPN analyzer has the correct gas
calibration. This can be accomplished by
introducing a known quantity of gas to the inlet and checking the response on
the instrument. The terminology for this
type of check is
Required Equipment: Precision flat head screwdriver
Reference: SOPMAN Pages XX-XX
Analyzer Setup
1.
The Background (BKG) and Coefficient
(COEF) on the analyzer should be set to 0.0 and 1.000 respectively.
2.
On the “S” model analyzers, use the
3.
On the “C” or “
Analyzer Response Check
1.
Open the Gas Calibration Worksheet and
enter the correct NO concentration (found on the gas cylinder) in the blank
provided.
2.
On DataView screen go to the Tools tab
and select Trigger Cals.
3.
In the Trigger Cals menu select NO3 Zero
and press Start Calibration.
4.
Under the tools tab select Query 1-Minute
Data.
5.
In the query data tool menu select the
Air
6.
Wait 7-10 minutes or until the data for
CH1, CH2 and CH3 are stable. Press Get
Data on the query data tool.
7.
Enter the Air
8.
Go to Trigger Cals and push Stop
Calibration then select NO3 NO
9.
Wait 7-10 minutes or until the data for
CH1, CH2 and CH3 are stable then press Get Data.
10.
Enter the Air
11.
The RF is now displayed at the far right
of the worksheet. If the RF is between
0.98 and 1.02, the calibration check is complete. Close the Trigger Cals and Query Data
windows.
12.
Save the Gas calibration worksheet then
print the TPN information and fax it to
13.
If the RF is not within spec continue
with the Analyzer Response Adjustment procedure.
Analyzer Response Adjustment
1.
Enter the as found data for NO3 Zero in
the as left section for NO3 Zero.
2.
On DataView go to the Tools tab and
select Trigger Cals.
3.
In the trigger cals menu select NO3 NO
4.
Under the tools tab select Query 1-Minute
Data.
5.
In the query data tool menu select the
Air
6.
Wait 7-10 minutes or until the data for
CH1, CH2 and CH3 are stable.
7.
Adjust the instrument to the Actual
Expected value from the As Found section of the worksheet.
a.
On “C” or “
b.
On “S” class analyzers open the front
door on the analyzer and adjust the PMT screw until channel one on the analyzer
or CH1 on DataView is approximately equal to the Actual Expected value from the
As Found section worksheet.
8.
When the analyzer adjustments are
complete, let the analyzer stabilize for 2-5 minutes.
9.
On the Query 1-Minute Data tool push Get
Data.
10.
Enter the Air
11.
The RF is now displayed at the far right
of the worksheet. If the RF is between
0.98 and 1.02, the calibration adjustment is complete. If the RF is not within spec, repeat the adjustment
part of the procedure again. (Note: In some cases CH3 will be higher than
1.02. Make sure that CH1 and CH2 are in
spec.)
12.
When the instrument is finished being
tested make sure to put it back into the run mode.
13.
Close the Trigger Cals and Query Data
windows.
14.
Save the Gas Calibration worksheet, and
then print the TPN information and fax it to