TN*/ TN Response
Check/ Adjustment
Purpose: The TN*/ TN response check/ adjustment is in
place to assure that the TN*/ TN analyzer has the correct gas calibration. This can be accomplished by introducing a
known quantity of gas to the inlet and checking the response on the
instrument. The terminology for this
type of check is
Required Equipment: Precision flat head screwdriver
Reference: SOPMAN Pages XX-XX
Analyzer Setup
1.
The Background
(BKG) and Coefficient (COEF) on the analyzer should be set to 0.0 and 1.000
respectively.
2.
On the “C” or “
1.
This procedure
will obtain the initial or current RF value for the TN*/TN analyzer.
2.
Open the Gas
Calibration Worksheet and enter the correct NO concentration (found on the gas
cylinder) in the blank provided.
3.
On DataView go to
the Tools tab and select Trigger Cals.
4.
In the trigger
cals menu select TN
5.
Under the tools
tab select Query 1-Minute Data.
6.
In the query data
tool menu select the Air
7.
Wait 10 minutes
for TN* and TN data to stabilize. Press
Get Data on the query data tool. If the
query data numbers are approximately equal to the number on DataView then the
data is stable and an average can be taken.
8.
Enter the Air
9.
Go to Trigger Cals
and push Stop Calibration then select TN NO
10.
Wait 10 minutes
for TN* and TN data to stabilize then press Get Data.
11.
Enter the Air
12.
The RF is now
displayed at the far right of the worksheet.
If the RF is between 0.98 and 1.02 the calibration check is
complete.
13.
Close the Trigger
Cals and Query Data windows.
14.
Save the Gas
calibration worksheet then print the
15.
If the RF is not
within spec continue with the Analyzer Response Adjustment procedure.
Analyzer Response Adjustment
1.
This procedure
will be used to adjust the analyzer and obtain a corrected RF value for the TN
analyzer.
2.
Enter the as
found data for TN
3.
On DataView go to
the Tools tab and select Trigger Cals.
4.
In the trigger
cals menu select TN NO
5.
Under the tools
tab select Query 1-Minute Data.
6.
In the query data
tool menu select the Air
7.
Wait 10 minutes
for TN* and TN data to stabilize. If the
query data numbers are approximately equal to the number on DataView then the
data is stable and an average can be taken.
8.
Adjust the
instrument to the Actual Expected value from the As Found section of the
worksheet.
9.
On “C” or “
10.
When the analyzer
adjustments are complete let the analyzer stabilize for 2-5 minutes.
11.
On the Query
1-Minute Data tool push Get Data.
12.
Enter the Air
13.
The RF is now
displayed at the far right of the worksheet.
If the RF is between 0.98 and 1.02 the calibration adjustment is
complete. If the RF is not within spec repeat
the adjustment part of the procedure again.
14.
When the
instrument is finished being tested make sure to put it back into the run mode.
15.
Close the Trigger
Cals and Query Data windows.
16.
Save the Gas Calibration
worksheet, and then print the