TN*/ TN Audit and Calibration
Purpose: The TN*/ TN response audit and calibration is
in place to assure that the TN*/ TN analyzer/system has the correct gas
response. This can be accomplished by
introducing a known quantity of gas to the inlet and auditing the response on
the instrument. The terminology for this
type of audit is
Required Equipment: Precision flat head
screwdriver
Analyzer Setup
1. The
Background (BKG) and Coefficient (COEF) on the analyzer should be set to 0.0
and 1.000.
2.
On the “C” or “
Analyzer Audit
1. This
procedure will obtain the current RF value for the TN*/TN analyzer.
3.
Open the Gas Calibration
Worksheet and enter the correct NO concentration (found on the gas cylinder) in
the blank provided.
4.
On Site Manager go to
the Tools tab and select Trigger Cals.
5.
In the Trigger Cals menu
select TN
6.
Under the Tools tab
select Query 1-Minute Data.
7.
In the Query Data Tool
menu select the Air
8.
Wait 13 minutes and
verify the Zeros for TN* and TN are stable (+/-0.1 ppb). Contact
9.
Enter the Air
10.
Go to Trigger Cals and
press Stop Calibration then select TN NO
11.
Wait 13 minutes and
verify the data for TN* and TN are stable (+/- 0.2ppb), then press Get Data.
12.
Enter the Air
13.
The RF is now displayed in
the worksheet. If the RF is between 0.98
and 1.02 the audit is complete.
14.
Close the Trigger Cals
and Query Data windows.
15.
Save the Gas calibration
worksheet.
16.
If the RF is not within
specification continue with the Analyzer Calibration procedure.
Analyzer Calibration
1. This
procedure will be used to adjust the analyzer and obtain a corrected RF value
for the TN analyzer.
2.
Enter the As Found data
for TN
3.
In Site Manager go to
the Tools tab and select Trigger Cals.
4.
In the Trigger Cals menu
select TN*/TN Zero and wait 5 minutes for data to be captured for First Light
use
5.
In the Trigger Cals menu
select TN NO
6.
Under the Tools tab
select Query 1-Minute Data.
7.
In the Query Data Tool
menu select the Air
8.
Wait 13 minutes and
verify the data for TN and TN* are stable (+/- 0.2 ppb). If the query data numbers are approximately
equal to the number on Site Manager, then the data is stable and an average can
be taken.
9.
Adjust the instrument to
the Actual Expected value from the As Found section of the worksheet.
10.
On “C” or “
11.
When the analyzer
calibration is complete let the analyzer stabilize for 5 minutes.
12.
On the Query 1-Minute
Data tool press Get Data.
13.
Enter the Air
14.
The RF is now displayed
in the worksheet. If the RF is between
0.98 and 1.02 the calibration is complete.
If the RF is not within specification repeat the calibration part of the
procedure again.
15.
When the instrument is
finished being calibrated make sure to put it back
into the Run mode.
16.
Close the Trigger Cals
and Query Data windows.
17.
Save the Gas Calibration
worksheet.
TN*/TN NO2 Conversion Efficiency Audit
1.
This procedure will
obtain the current RF value for the NO2 Conversion Efficiency.
2.
Open the Gas Calibration
Worksheet and go to the NO2 Conv Eff Worksheet tab. Enter the correct NO and NO2 concentrations
(found on the gas cylinder) in the blanks provided.
3.
An NO audit of the
analyzer has to be completed before the NO2 conversion efficiency can be
completed. You may use that day’s
information if available. If it is available
enter the information and start with Step 15.
Otherwise continue with the next Step.
4.
On the Site Manager
screen go to the Tools tab and select Trigger Cals.
5.
In the Trigger Cals menu
select TN Zero and press Start Calibration.
6.
Under the Tools tab select
Query 1-Minute Data.
7.
In the query data tool
menu select the Air MFC, Gas
8.
Wait 13 minutes and
verify the Zeros for TN* and TN are stable (+/-0.1 ppb). Contact
9.
Enter the Air
10. Go to Trigger Cals and press Stop Calibration
then select TN*/TN NO GRC Point 1 and press Start Calibration.
11. Wait 13 minutes and verify the data for TN* and
TN are stable (+/- 0.2ppb), then press Get Data.
12. Enter the Air MFC, Gas
13. The RF is now displayed at the far right of the
worksheet.
14. Save the Gas Calibration Worksheet.
15. In the Trigger Cals menu select TN Zero and
press Start Calibration.
16. Under the Tools tab select Query 1-Minute Data.
17. In the query data tool menu select the Air MFC,
Gas
18. Wait 13 minutes and verify the Zeros for TN* and
TN are stable (+/-0.1 ppb). Contact
19. Enter the Air
20. Go to Trigger Cals and press Stop Calibration
then select TN*/TN NO2 GRC Point 1 and press Start Calibration.
21. Wait 13 minutes and verify the data for TN* and
TN are stable (+/- 0.2ppb), then press Get Data.
22. Enter the Air MFC, Gas
23. The RF is now displayed at the far right of the
worksheet.
24. If the RF is less than 0.95 it is out of spec
and the RF will turn red.
25. Contact
26. Close
the Trigger Cals and Query Data windows.
27. Save
the Gas Calibration Worksheet.