CO
Trap Efficiency Check
Purpose:
The purpose of this procedure is to check the efficiency of the CO Trap.
Required Equipment: None
Reference: SOPMAN
Data Collection
1.
On
DataView, go to the Setup tab and select Process Control.
2.
On
the lower left corner of the Process Control screen check the box next to
Freeze Process Programs.
3.
In
the left column of Process Controls, highlight CO/SO2 Zero from the list and then
select Turn ON.
4.
Under
the Tools tab, select Query 1-Minute Data. In the query data tool menu select
CO.
5.
Wait
5-10 minutes or until the CO number is stable.
A stable number has been reached when the query data numbers are
approximately equal to the number on DataView.
6.
Record
the CO value for the last one minute average and label it GRC Zero
7.
Select
Turn OFF in the Process Control program.
8.
Highlight
CO/SO2 GRC Point 1 from the list, and then select Turn On.
9.
Wait
until the CO value is stable.
10.
Record
the CO value for the last one minute average and label it Point1.
11.
Select
Turn OFF in the Process Control program.
12.
13.
14.
Wait
until the CO value is stable.
15.
Record
the CO value for the last one minute average and label it GRC Zero with Filter
Zero.
16.
Highlight
CO/SO2 Zero from the list, then select Turn OFF. Leave the CO Zero running.
17.
Highlight
CO/SO2 GRC Point 1 from the list; select Turn ON. Now both the CO Zero and CO/SO2
GRC Point 1 are running.
18.
Wait
until the CO value is stable.
19.
Record
the CO value for the last one minute average and label it Point1 with Filter
Zero.
20.
Turn
OFF both the CO zero and the CO/SO2 GRC Point 1.
21.
Close
the Process Control program.
Calculating the CO Trap
Efficiency
1.
To calculate the CO Trap Efficiency four
pieces of information are needed (GRC Zero, Point1, GRC
Zero with Filter Zero and Point 1 with Filter Zero).
2.
The
formula for calculating the CO Trap Efficiency is as follows:
Efficiency = 1 – [(Point1 with
Filter Zero – GRC Zero with Filter Zero) / (Point1 – GRC Zero)]
(Example: Eff. = (1 – [(250 –
200) / (1200 – 200)]) x 100) = 95%
3.
The
CO Trap Efficiency should be 100 % +/-0.01.