5400 Run Definition Check Procedure

 

Purpose:  Follow the steps below to check the run definition setup screen on the 5400 instrument.

 

Required Equipment:  None

 

Reference:  SOPMAN

 

1.        This check is completed with the instrument Stop mode.

2.        From the Main Screen press the F3/ Run Def button.

3.        Use the edit button and input the correct setting if found to be different than listed in chart below.

4.        Enter must be pressed before any change will be accepted by the CPU.

 

ARA SEARCH 5400 Run Definitons:

Collection Time:   1 Hour                    Analysis Type:     Reg

Col Temp:              0.0 C                       P1:           0.0 C                       D1:          0.0 sec

Aburn Temp:        0.0 C                       P2:           0.0 C                       D2:          0.0 sec

Ext. Temp:              0.0 C                       P3:           0.0 C                       D3:          0.0 sec

Col Flow:               16.7                         P4:           600 C                      D4:          240 sec